108 年 國立成功大學化學研究所《分析化學》
第 1 題5 分
Which of the following statements is correct?
(A) TD glassware is calibrated to deliver a known volume of liquid at a given temperature that is printed on the glassware. However, TC glassware is calibrated to contain a known volume at a given temperature that is printed on the glassware.
(B) Precision describes the reproducibility of measurements or the closeness of the measurement to the true or accepted value. It can be expressed by standard deviation, variance and coefficient of variation.
(C) Systematic errors (determinate error) have assignable causes and definite values. They are often unidirectional. Types of these errors include: instrumental errors, method errors and personal errors.
(D) Constant errors are independent of the size of the sample being analyzed. The effect of a constant error becomes less serious as the size of the quantity measured decreases.
(E) In the feedback system flow diagram, the difference between the two states is used to change a controllable quantity that results in a change in the state of the system. The process provides continuous monitoring and feedback to maintain the controllable quantity.
登入後即可作答並保存紀錄。
核心觀念
-
TD (To‑Deliver) 與 TC (To‑Contain) 玻璃器皿的校正方式
- TD:校正目的是在特定溫度下送出(deliver)已知體積的液體;使用者在操作時必須自行把握容器內的液體量,器皿本身保證的是「送出」的體積。常見於容量吸管、滴定管、移液管。
- TC:校正目的是在特定溫度下容納(contain)已知體積的液體;器皿本身保證的是容器內裝滿的體積,使用者只需把液體倒入器皿即可得到該體積。常見於容量瓶、量筒、滴定瓶。
-
精密度(Precision) vs 正確度(Accuracy)
- 精密度:同一樣本重複測量結果之再現性,與真值的接近程度無關。常用 標準差 、變異數 、變異係數 表示。
- 正確度:測量結果與真值(或公認值)的接近程度,通常以誤差 (bias) 來描述。
-
系統誤差(Systematic error)
- 來源可辨識(儀器、方法、校正程序),且在同樣條件下會產生固定方向的偏差,屬決定性(deterministic)誤差。
- 個人操作的隨機偏差屬隨機誤差,不應列入系統誤差。
-
恆定誤差(Constant error)
- 絕對誤差大小與測量量的大小無關,相對誤差會隨測量量增大而減小。因此「量測值越大,恆定誤差相對影響越小」是正確描述。
-
回饋控制系統(Feedback system)
- 以差值(error signal)驅動控制元件改變可控變數(manipulated variable),進而調整系統狀態(process variable),形成閉環。此過程需要持續監測與即時回饋。
解題方法
本題為單選題,須判斷每一敘述的正確性。
- 先根據教材概念檢視每個選項的關鍵詞與定義,確認其是否符合標準定義。
- 若選項中出現與教材不符或概念混淆的敘述,即判定為錯誤。
- 只要找到唯一符合所有概念的敘述,即為正確答案。
選項分析
| 選項 | 判斷 | 說明 |
|---|---|---|
| (A) | ✅ 正確 | TD 器具的校正目的是送出(deliver)已知體積,TC 器具的校正目的是容納(contain)已知體積,均在標示的校正溫度(通常 20 °C)下完成,敘述與教材完全一致。 |
| (B) | ❌ 錯誤 | 精密度僅描述再現性,不涉及與真值的接近程度;後半段「或與真實值的接近」屬於正確度的定義,混淆兩者概念。 |
第 2 題5 分
Which of the following methods can be used to measure the thickness of thin films?
(A) Tunneling electron microscopy
(B) Spectroscopic ellipsometry
(C) Electrospray ionization mass spectrometry
(D) Atomic force microscopy
(E) Atomic X-ray spectrometry
登入後即可作答並保存紀錄。
核心觀念
薄膜(thin film)厚度的量測方法主要分為光學干涉/偏振法、探針表面輪廓/掃描顯微法與截面電子顯微法等。
選項分析
- (A) Tunneling electron microscopy:一般並無「Tunneling electron microscopy」的標準術語,常見為掃描穿隧顯微鏡(Scanning Tunneling Microscopy, STM)或穿透式電子顯微鏡(Transmission Electron Microscopy, TEM)。STM 主要量測導電表面的電子態密度與原子級表面形貌,無法直接量測一般薄膜厚度;TEM 則需製備截面樣品(cross-sectional TEM)才能量測厚度,選項名稱本身亦非標準厚度量測技術。
- (B) Spectroscopic ellipsometry(光譜橢圓偏振儀):為量測薄膜厚度與光學常數(折射率 、消光係數 )最常用且精確的非破壞性光學技術,可量測從亞奈米至數微米的單層或多層薄膜厚度。
- (C) Electrospray ionization mass spectrometry(電灑游離質譜法,ESI-MS):用於測定溶液中大分子或極性化合物的分子量與結構,與薄膜厚度量測無關。
第 3 題5 分
Which of the following statements related to Raman vs. Infrared is correct?
(A) For a given bond, the energy shifts observed in a Raman experiment should be different to the energies of its infrared absorption bands.
(B) The molecule detected in Raman and Infrared need possess a permanent dipole moment.
(C) Water can be used as a solvent in Raman measurement, but water cannot be used due to its intense absorption of Infrared region.
(D) The vibration is Raman active if it causes a change in dipole moment, and vibration is Infrared active if there is change in polarizability.
(E) Raman is due to the scattering of light by the vibrating molecules, and Infrared is the result of absorption of light by vibrating molecules.
登入後即可作答並保存紀錄。
核心觀念
-
紅外 (IR) 吸收
- 原理:分子振動能夠吸收與其振動頻率相同的電磁波。
- 必要條件:該振動必須伴隨 偶極矩 (dipole moment) 的變化。
- 只要振動導致 ( 為偶極矩, 為正常座標),即為 IR‑active。
-
拉曼 (Raman) 散射
- 原理:入射光子被分子散射,散射光子能量可因分子振動而改變(Stokes / anti‑Stokes)
- 必要條件:該振動必須伴隨 分子極化率 (polarizability) 的變化。
- 只要振動導致 ( 為極化率張量),即為 Raman‑active。
-
能量位移
- 無論是 IR 吸收還是 Raman 散射,觀測到的 振動能量(頻率 )皆由分子本身的振動模式決定,理論上兩者應相同。Raman 只測量相對於入射光的位移(Raman shift),而 IR 直接測量吸收頻率。
-
溶劑效應
- 水在 IR 区段(尤其是 3000–3700 cm⁻¹、1600 cm⁻¹)有強烈的吸收,會遮蔽樣品訊號。
- 在 Raman 測量中,水的 Raman 效應相對較弱,且可使用特殊濾光或近紅外激發波長降低背景,故常被選作溶劑。
解題方法
針對每一選項,檢視其敘述是否符合上述三個核心概念(偶極矩 vs. 極化率、能量等價、溶劑特性)。只要有一項與基本原理相悖,即為錯誤選項;全部符合則為正確選項。
選項逐一分析
| 選項 | 內容 | 判斷 | 理由 |
|---|---|---|---|
| A | 「對同一鍵,Raman 實驗觀測的能量位移應與其 IR 吸收帶的能量不同。」 | ❌ 錯誤 | 同一振動模式的能量(頻率)在兩種光譜中相同。Raman 只呈現Raman shift(相對於入射波長),其數值等於振動頻率 ,與 IR 吸收頻率一致。 |
| B | 「Raman 與 IR 所偵測的分子皆須具永久偶極矩。」 | ❌ 錯誤 | IR 需要偶極矩變化,但 Raman 不需要。即使分子無永久偶極(如 、),只要極化率隨振動改變,即可在 Raman 中觀測。 |
| C | 「水可用於 Raman 測量,但因其在 IR 區段吸收強烈,水不可用於 IR 測量。」 | ✅ 正確 | 水的 O–H 伸縮振動在 3000–3700 cm⁻¹ 具有極強的 IR 吸收,會遮蔽樣品訊號;而在 Raman 中,水的 Raman 效應相對較弱,且可透過選擇激光波長或使用去除背景的技術抑制。因此此敘述符合實務觀測。 |
| D | 「若振動導致偶極矩變化則為 Raman active;若導致極化率變化則為 IR active。」 | ❌ 錯誤 | 完全顛倒了判準。 |
第 4 題5 分
Which of the following terms used to describe the theory of band broadening in chromatography?
(A) The single and multiple path term, A.
(B) The longitudinal and lateral diffusion term, B/u.
(C) The stationary-phase mass-transfer term, Csu.
(D) The mobile-phase mass-transfer term, Cmu.
(E) All above options are correct.
登入後即可作答並保存紀錄。
核心觀念
色譜法的譜帶展寬(band broadening)速率理論以 van Deemter 方程式描述,塔板高度 為各展寬來源之和:
其中 為移動相線流速。四項分別為:
- :多路徑(渦流擴散)項,溶質分子經過填充顆粒間長短不一的路徑造成展寬,與流速無關。
- :縱向擴散項,溶質沿管柱軸向順濃度梯度擴散,流速愈低愈明顯。
- :固定相質傳項,溶質進出固定相所需時間造成的展寬,隨流速增加。
- :移動相質傳項,溶質在移動相內擴散到固定相表面所需時間造成的展寬,隨流速增加。
選項分析
- (A): 項即多路徑項,屬於展寬理論的組成,題目中 single 字樣只是對該項的描述用語,列為展寬項是對的。
- (B): 為縱向(軸向)擴散項,是展寬理論的組成;題目的措辭略為寬鬆,但所指的就是 這一項。
第 5 題5 分
Which of the following statements related to ionization in mass spectrometry is correct?
(A) Ions formation by electron ionization (EI) or electrospray ionization (ESI) must be in a vacuum.
(B) Matrix-assisted laser desorption/ionization (MALDI) enables to generate significant fragmentation, which is very informative about structure of a molecule.
(C) Fast atom bombardment (FAB) is an ionization technique used in mass spectrometry in which a beam of high speed atoms with low energy strikes a sample surface to create ions.
(D) Sample for ESI must be dissolved in an appropriate solvent, but sample for MALDI need be applied in an appropriate matrix.
(E) Chemical ionization (CI) generates much less fragments than with EI; it is more likely to observe an ion closely related to the original molecule.
登入後即可作答並保存紀錄。
核心觀念
質譜法常見的游離源可依操作壓力與軟硬程度區分:
- 大氣壓游離(API):如電噴霧游離(ESI)、大氣壓化學游離(APCI),在常壓下產生離子後再導入高真空質量分析器。
- 真空游離:如電子游離(EI)、化學游離(CI,低壓反應氣體)、快原子撞擊(FAB)、基質輔助雷射脫附游離(MALDI,多數在真空下進行)。
- 軟游離(soft ionization):如 ESI、MALDI、CI,碎片少、易見準分子離子;硬游離(hard ionization):如 EI,碎片豐富。
選項分析
- (A) 錯誤:電子游離(EI)必須在高真空下進行,但電噴霧游離(ESI)是在大氣壓下產生離子,之後才經由介面導入真空中的質量分析器,並非兩者都必須在真空中形成離子。
- (B) 錯誤:基質輔助雷射脫附游離(MALDI)屬於軟游離技術,主要產生完整的單電荷準分子離子(如 ),幾乎不產生大量碎片;若需結構碎片資訊,須搭配串聯質譜(MS/MS)。
第 6 題5 分
Which of the following methods can provide elemental and chemical information?
(A) Scanning tunneling microscopy
(B) Energy-dispersive X-ray spectroscopy
(C) Dynamic light scattering
(D) Inductively coupled plasma atomic emission spectroscopy
(E) X-ray photoelectron spectroscopy
登入後即可作答並保存紀錄。
核心觀念
- 元素資訊:指能辨識樣品中存有何種原子(元素),常透過特徵能量(X‑ray、光子、質子等)或散射角度取得。
- 化學資訊:指提供元素的化學狀態(氧化態、鍵結環境、配位數、分子結構等),通常需要測量 鍵合能、光電子結合能、化學位移 等資訊。
- 常見分析技術依據偵測之訊號類型分為:
- 形貌/結構觀察(STM、DLS)
- 光譜分析(EDS、ICP‑AES、XPS)
解題方法
- 先釐清每個方法的偵測原理與可得到的訊號。
- 判斷該訊號是否能同時告訴我們「是哪個元素」以及「該元素在什麼化學環境」:
- 若僅是 特徵能量(如 X‑ray 線),只能給出元素身份。
- 若偵測到 化學位移或結合能變化,則能提供化學資訊。
- 逐一比對選項,找出同時具備上述兩項能力的技術。
選項分析
| 選項 | 方法說明 | 能否提供 元素資訊 | 能否提供 化學資訊 | 結論 |
|---|---|---|---|---|
| (A) Scanning tunneling microscopy (STM) | 以探針量測樣品表面電子隧穿電流,得到原子尺度的表面拓撲影像。 | 只能依賴表面電導差異間接推測,無法直接辨識元素。 | 只能觀測表面形貌與電子態密度分佈,無化學鍵結資訊。 | ❌ |
| (B) Energy‑dispersive X‑ray spectroscopy (EDS) | 受激發樣品發出特徵 X‑ray,根據 X‑ray 能量判斷元素。 | ✅ 直接給出元素的特徵 X‑ray 能量。 | 只能得到元素的相對濃度,無法分辨氧化態或配位環境。 | ❌(僅元素) |
| (C) Dynamic light scattering (DLS) | 依粒子在懸浮液中的布朗運動散射光的變化,算出粒徑分布。 | 與光散射強度無關,無法辨識元素。 | 完全不提供化學資訊。 | ❌ |
| (D) Inductively coupled plasma atomic emission spectroscopy (ICP‑AES) | 樣品在等離子體中被激發,原子返回基態時發射特徵光譜。 | ✅ 透過發射波長辨識元素並量測濃度。 |
第 7 題5 分
Which of the following statements is correct?
(A) In titration, the titrant may be a standard solution of a known chemical or an electric current of known magnitude.
(B) The end point is the point in a titration when the amount of added standard reagent is equivalent to the amount of analyte.
(C) Back-titrations are often required when the rate of reaction between the analyte and reagent is slow or when the standard solution lacks stability.
(D) There are three typical types of titration curves: the sigmoidal curve, the linear segment curve, and the Conchoid of de Sluze curve.
(E) In particulate gravimetry, the analyte can be determined by separating it from the matrix of sample using a filtration or an extraction.
登入後即可作答並保存紀錄。
核心觀念
本題考查滴定分析與重量分析的基本定義:
- 滴定劑(titrant):濃度已知的標準溶液,或在庫侖滴定中由已知電流通過一定時間所產生的定量試劑。
- 當量點(equivalence point):加入滴定劑的量,恰好符合分析物化學計量需求的點。
- 終點(end point):由指示劑顏色、電位、電流等實驗訊號判定滴定完成的觀察點。
- 返滴定(back titration):先加入過量且已知量的標準試劑,使其與分析物反應,再以另一標準溶液滴定剩餘試劑。
- 顆粒重量分析(particulate gravimetry):將分析物或含分析物的顆粒與樣品基質分離,再測量其質量。
解題方法
逐一比對各選項與上述定義。特別注意 當量點與終點不可混為一談;此外,本題有多個正確選項,應採複選方式判斷。
選項分析
(A)正確
在一般滴定中,滴定劑通常是濃度已知的標準溶液。
在庫侖滴定(coulometric titration)中,滴定試劑可由電解反應在原位產生,其量由通過電解槽的電量決定:
其中 為電流, 為通電時間。依照法拉第定律,產生的試劑莫耳數為:
其中 為電子轉移數, 為法拉第常數。因此,已知電流並測量通電時間,即可精確決定滴定劑的量。
所以本選項所述概念正確。
(B)錯誤
題目將「終點」的定義寫成「加入標準試劑的量與分析物等量的點」,這其實是當量點的定義。
- 當量點:依化學計量關係,滴定劑與分析物恰好完全反應的理論點。
- 終點:藉由實驗訊號觀察到滴定完成的實際判定點。
理想情況下,終點應盡量接近當量點,但兩者並不完全相同,兩者的差異稱為滴定誤差。
因此,本選項錯在把 endpoint 說成 equivalence point。
(C)正確
返滴定適用於下列情況:
- 分析物與試劑的反應速率太慢,無法直接滴定。
- 分析物不易直接滴定,或缺乏合適的指示劑。
- 標準試劑穩定性不足,難以直接作為準確滴定劑使用。
返滴定的計算概念為:
例如先加入已知莫耳數的試劑 :
再依反應計量關係求得分析物含量。因此,本選項正確。
(D)正確
第 8 題5 分
Which of the following statements is correct?
(A) When the gas diffuses across the permeable membrane of a gas-sensing electrode, it alters the composition of the inner solution, which is monitored with an ion-selective electrode.
(B) Potentiometric electrodes are designed to respond to molecules by using a chemical reaction that produces an ion whose concentration can be determined using a traditional ion-selective electrode.
(C) Polarography is a voltammetric technique that uses a mercury electrode and an unstirred solution.
(D) Amperometry is a potentiometric method in which applies an alternating potential difference to the electrode and measures the resulting current.
(E) Coulometric methods are based on Faraday's law that the partial charge or current passed during an electrolysis is inversely proportional to the amount of reactants and products in the redox reaction.
登入後即可作答並保存紀錄。
核心觀念
電化學分析法依量測訊號可分為:
- 電位法(Potentiometry):在無電流()條件下量測電極電位,如離子選擇性電極(ISE)、氣體感測電極與電位式生物感測器。
- 伏安法(Voltammetry)與極譜法(Polarography):施加掃描電位並量測電流;極譜法特指使用汞電極(如滴汞電極 DME、靜汞滴電極 SMDE)於靜止溶液中的伏安法。
- 安培法(Amperometry):施加固定電位並量測電流隨時間或濃度的變化。
- 庫侖法(Coulometry):依據法拉第電解定律,量測完全電解所需的總電量 。
選項分析
- (A) 正確:氣體感測電極(gas-sensing electrode)的原理是待測氣體擴散穿透氣體滲透膜進入內部溶液並改變其離子組成(如 pH 值),再由內部的離子選擇性電極(ISE)監測該離子濃度變化。
- (B) 正確:針對分子設計的電位式電極(如酵素電極等電位式生物感測器),可利用化學/酵素反應將目標分子轉化為特定離子,再以傳統的離子選擇性電極測定其濃度。
- (C) 正確:極譜法(polarography)屬於伏安法的一種,使用汞電極(如 DME 或 SMDE)並在未攪拌的靜止溶液(unstirred solution)中進行量測,使質量傳遞僅受擴散控制。
第 II.1 題10 分
Salting out is an effect which is used as a method for protein purification in the laboratory. Salting out usually relies on changing the protein solubility by adding lots amount of salts (electrolytes) in to a protein solution. At higher salt concentration, protein solubility usually decreases, leading to protein precipitation. Please explain the principle of salting-out effect and the reason why proteins precipitate at higher salt concentration.
登入後即可作答並保存紀錄。
此題為問題解決與簡答題,主要考察考生對蛋白質純化技術中「鹽析 (salting out)」原理的理解。
核心觀念:
鹽析是一種利用改變溶液離子強度來降低蛋白質溶解度,使其沉澱析出的蛋白質純化方法。其原理與蛋白質分子表面的電荷、水合作用以及離子與水分子之間的競爭有關。
原理闡述:
- 蛋白質的溶解度與水合作用: 蛋白質分子表面帶有電荷,並且與水分子形成水合層,這層水合層有助於蛋白質溶解在水中。蛋白質的溶解度取決於其表面電荷與水分子相互作用的強度,以及蛋白質分子之間的相互作用。
- 鹽的加入與離子強度: 當向蛋白質溶液中加入大量鹽(如硫酸銨 (NH₄)₂SO₄、氯化鈉 NaCl 等)時,鹽會解離成離子。這些離子會強烈地與水分子發生離子-偶極相互作用,形成離子水合層。
- 離子與蛋白質的競爭: 鹽離子對水分子的強烈親和力,會與蛋白質爭奪溶液中的水分子。在高鹽濃度下,水分子被鹽離子「搶奪」走,導致蛋白質表面的水合層變薄,蛋白質分子之間的相互作用(疏水作用和靜電作用)相對增強。
- 蛋白質沉澱: 隨著水合作用的減弱和蛋白質分子間相互作用的增強,蛋白質分子更容易聚集在一起,形成蛋白質聚集體,最終導致蛋白質溶解度下降並從溶液中沉澱析出。這個過程稱為「鹽析」。
蛋白質在高鹽濃度下沉澱的原因:
在高鹽濃度下,加入的無機鹽(電解質)會離解成大量的陽離子和陰離子。這些離子與水分子之間存在很強的離子-偶極作用,形成穩定的水合離子。這些水合離子會優先與水分子結合,從而「奪走」了水分子,削弱了蛋白質分子表面的水合作用。
蛋白質的溶解度很大程度上依賴於其表面的水合層,水合層可以防止蛋白質分子之間直接接觸並聚集。
第 II.2 題10 分
The quantitative determination of mercury in all types of samples is vital important, due to the toxicity and widespread distribution of organic mercury compounds in the environment. Atomic absorption spectrometry (AAS) has been widely used to determine the concentration of elements in analytical samples. Please describe and explain the principle of the AAS-based method for determination of mercury concentration in samples.
登入後即可作答並保存紀錄。
此題為問題解決與簡答題,考察考生對原子吸收光譜法 (AAS) 在汞 (Hg) 元素定量分析中的應用原理的理解。
核心觀念:
原子吸收光譜法 (AAS) 的基本原理是基於基態原子對特定波長光的吸收。不同元素的原子在吸收特定波長的光後會躍遷到激發態,其吸收光譜具有元素特異性。通過測量待測元素在基態原子蒸氣狀態下對其特徵吸收波長的吸收程度,可以定量分析該元素的濃度。
AAS 測量汞的原理:
-
樣品處理與原子化:
- 樣品類型: 汞在環境樣品(水、土壤、生物組織等)中可能以不同形態存在(如無機汞、有機汞)。為了進行 AAS 分析,通常需要將樣品轉化為易於原子化的形態。
- 還原蒸發法 (Vapor Generation / Cold Vapor AAS, CVAAS): 汞是唯一一種在室溫下即可呈氣態的液態金屬。這是汞在 AAS 分析中的一個顯著優勢。對於含有汞的樣品(無論是無機汞還是可以轉化為無機汞的有機汞),通常會加入強還原劑,如硼氫化鈉 (NaBH₄) 或氯化亞錫 (SnCl₂),在酸性條件下將所有形態的汞還原為單質汞 (Hg⁰)。
- 產生汞蒸氣: 還原產生的汞蒸氣 (Hg⁰) 會被導入一個氣流系統(如氣體傳輸管或流動池),形成一個富含汞原子的「原子蒸氣」。
- 原子化: 在 CVAAS 中,汞蒸氣本身就處於原子狀態,不需要額外的加熱原子化步驟(如火焰或石墨爐)。這與其他元素的 AAS 分析不同。
-
光源與吸收:
- 光源: 使用一個能夠發射汞特徵吸收波長光的空心陰極燈 (Hollow Cathode Lamp, HCL) 或其他汞放電燈作為光源。汞的特徵吸收波長為 253.7 nm。
- 吸收: 當 253.7 nm 的光通過含有汞蒸氣的流動池時,基態的汞原子會吸收此波長的光,發生電子躍遷。
-
信號檢測與定量:
- 檢測器: 光譜儀中的檢測器(如光電倍增管 PMT)測量通過樣品池的光強度。
- 吸收測量: 通過比較有樣品蒸氣存在時和不存在時的光強度,可以計算出汞蒸氣對 253.7 nm 光的吸收度 (Absorbance)。
- 定量: 根據朗伯-比爾定律 (Lambert-Beer Law),吸收度與樣品中汞原子的濃度成正比。通過建立一系列已知濃度的汞標準溶液,進行還原蒸發處理,測量其吸收度,繪製標準曲線。然後,將待測樣品進行相同的處理,測量其吸收度,並從標準曲線上讀出對應的汞濃度。
優勢與注意事項:
- 高靈敏度: CVAAS 對汞的檢測極為靈敏,檢測限可達 ng/L (ppt) 級別,非常適合環境監測。
第 II.3 題10 分
For the titration reaction A + B → C, where A is the analyte, B is the titrant, and C is the product, the titration is carried out via a UV-Vis spectrophotometer. The end point of titration is detected at 550 nm, based on the absorbance information below. Please explain and draw the titration curve at 550 nm.
| Substance | Wavelengths absorbed (nm) |
|---|---|
| A | 400 |
| B | <400, 500~700 |
| C | <400 |
登入後即可作答並保存紀錄。
核心觀念
- 光吸收與濃度的關係:在單波長測定中,吸光度 與溶液中所有在此波長有吸收的物種之濃度呈線性關係(比爾‑朗伯定律)
其中 為摩爾吸光係數, 為光徑長, 為第 種物質的摩爾濃度。 - 滴定的化學計量:本題假設 與 以 1:1 摩爾比反應生成 ,即
當加入 体积的滴定劑 (濃度 )時, 的剩餘摩爾數為 ,而 的剩餘摩爾數為 (若 超過等當點)。 - 波長選擇的意義:在本題 550 nm 處,根據表格
- 吸收:400–600 nm → 吸收
- 吸收:500–700 nm → 吸收
- 吸收:<400 nm → 不吸收
因此測量的吸光度是 與 的貢獻之和, 不參與。
解題方法
-
寫出吸光度表達式(以體積 為自變數)
- 當 (等當點之前)
其中 ,, 為混合後的總體積。 - 當 (等當點之後)
其中 。
- 當 (等當點之前)
-
判斷曲線形狀
- 若 (本題未給定,常假設相近),則等當點前 的減少與 的增加大致相抵,吸光度呈緩慢下降或近似水平。
- 等當點後,只剩過量的 ,吸光度開始上升(若 )或保持不變(若滴加的 吸光度極小)。
- 若 ,等當點前吸光度會顯著下降,等當點後則轉變為緩慢上升。
- 由於 完全不吸收,等當點前的 逐漸被“隱形”化合物 取代,導致吸光度下降是本題的關鍵指示。
-
繪製滴定曲線
- X 軸:滴定劑 加入的體積(或等效當量)
- Y 軸:550 nm 處的吸光度
- 曲線分段:
- 起始點()吸光度最高,僅有 的貢獻。
- 隨著 增加,吸光度呈下降斜率(取決於 )。
- 在等當點 處出現拐點,之後吸光度不再下降,若 可測則呈輕微上升;若 很小則趨於水平平台。
- 示意圖(文字版)
第 II.4 題10 分
Ethylenediaminetetraacetic acid (EDTA) is widely used in complexometric titrations and analysis of water hardness. Water hardness is determined by titrating Ca2+ and Mg2+ ions with a EDTA solution at a buffered pH of 10. Please explain the following questions.
(A) Why is the water sample buffered to the pH of 10 during water hardness analysis?
(B) What problems might you expect at a higher pH or a lower pH of the water sample?
登入後即可作答並保存紀錄。
核心觀念
以 EDTA 錯合滴定測定水硬度( 與 總量)時,系統受三個平衡共同限制:
- EDTA 酸鹼解離平衡:僅完全去質子化的 具有最強錯合力,條件生成常數為 。
- 金屬離子水解/沉澱平衡:高 pH 下 易生成 沉澱。
- 金屬指示劑(鉻黑 T,Eriochrome Black T, EBT)之酸鹼與錯合變色平衡:
- :游離 EBT()為酒紅色。
- :游離 EBT()為藍色。
- :游離 EBT()為橙色。
- 金屬-EBT 錯合物()為酒紅色。
解題方法
-
(A) 緩衝於 的原因:
- 確保反應完全度:在 時,EDTA 主要以 與 存在(),使 與 的條件生成常數 足夠大,滴定終點突躍明顯。
- 維持指示劑有效變色: 落在 EBT 呈藍色()的區間();滴定前樣品呈酒紅色(),到達終點時 EDTA 搶走 釋出游離 ,溶液由酒紅色變為純藍色,對比極為清晰。
- 兼顧避免氫氧化物沉澱: 尚未高到使大量 立即沉澱為 ,且 緩衝液亦有助於防止金屬離子沉澱。
-
(B) pH 過高或過低可能產生的問題:
- pH 過低(例如 ,尤其 ):
- EDTA 大量質子化使 急劇下降, 變小(特別是 - 的 較小,滴定不完全、終點模糊)。
- 當 時,游離 EBT()本身即呈酒紅色,與 同色,到達終點時無法觀察到顏色變化。
- pH 過高(例如 ):
- pH 過低(例如 ,尤其 ):
第 II.5 題10 分
Micro total analysis systems (µTAS), or so-called laboratory-on-a-chip (LOC), are the devices that automate and include all necessary steps for chemical analysis, such as sampling, sample transport, chemical reactions, separation and detection.
(A) What is the advantages of the µTAS or LOC, compared to traditional chemical analysis methods?
(B) Please describe the basic principle of separation approaches used in µTAS or LOC.
登入後即可作答並保存紀錄。
核心觀念
微全分析系統(micro total analysis systems, µTAS)或晶片實驗室(lab-on-a-chip, LOC)是利用微機電/軟微影製程,將採樣、前處理、混合反應、分離與偵測整合於微米級流道晶片上的微型化分析平台。
解題方法
-
(A) µTAS / LOC 相較於傳統分析方法的優勢:
- 樣品與試劑消耗極少:微流道體積僅奈升()至微升()等級,大幅降低昂貴試劑與珍貴樣品用量,並減少廢液。
- 分析與反應速度極快:微尺度下擴散距離極短(擴散時間 )、熱傳導效率高,反應與分離可在數秒至數分鐘內完成。
- 高度整合與自動化:將採樣、反應、分離、偵測整合於單一晶片,減少人為操作誤差並提升再現性。
- 高通量與平行化:可在同一晶片設計多條平行微流道,同時處理大量樣品。
- 可攜性與現場/床邊檢測(POCT):體積小、耗能低,適合開發為手持式設備應用於環境監測與臨床即時診斷。
- 安全性高:密閉且微量的流體系統可降低有毒、具傳染性或易爆試劑的操作風險。
-
(B) µTAS / LOC 常見分離技術的基本原理:
- 電泳分離(Electrophoretic separation,如微晶片毛細管電泳 MCE):
- 施加上游至下游的電場後,帶電分析物依荷質比與電泳遷移率(electrophoretic mobility)差異而以不同速度遷移;同時微流道壁面雙電層在電場下產生電滲流(electroosmotic flow, EOF),可作為無需機械幫浦的平坦流場推動液體,大幅降低譜帶展寬。
- 層析分離(Chromatographic separation,如 、 與毛細管電層析 CEC):
- 依各組分在動相與靜相(塗覆於微流道壁面或微柱填充顆粒)之間的分配係數差異達成滯留時間分離;若以電滲流取代高壓幫浦驅動動相,即為毛細管電層析(CEC)。
- 擴散分離(Diffusion-based separation,如 H-filter):
- 微流道中雷諾數極低,流體呈穩定層流(laminar flow)而不發生紊流混合;兩股液流並流時,僅小分子能快速橫向擴散進入另一股流體,大分子或微粒則因擴散係數小而留在原流道,藉此實現連續分離。
- 過濾與物理場分離(Filtration / Centrifugal / Dielectrophoretic separation):
- 電泳分離(Electrophoretic separation,如微晶片毛細管電泳 MCE):
第 II.6 題10 分
Please explain the reason why compound A exhibits stronger fluorescence than compound B in acetonitrile.
[Image of two chemical structures, labeled A and B]
登入後即可作答並保存紀錄。
核心觀念
本題評量儀器分析中分子螢光光譜法(Molecular Fluorescence Spectrometry)的基本原理,核心在於影響螢光量子產率(fluorescence quantum yield, )的分子結構與環境因素:
- 螢光量子產率公式: 其中 為螢光輻射放光的速率常數; 為非輻射去活化(non-radiative deactivation)速率常數總和,包含內部轉換(internal conversion, )、外部轉換(external conversion, )、系統間穿隧(intersystem crossing, )及預解離/解離(predissociation, )。
- 分子結構剛性(Structural Rigidity)與共平面性(Coplanarity):
分子結構越剛硬、平面性越佳,分子內的振動與轉動自由度越低,內部轉換速率常數 越小,受激分子越傾向以輻射形式放光,因而大幅提升螢光強度與量子產率。 - 取代基效應(Substituent Effect)與分子內電荷轉移(ICT):
引入強給電子基團(如 )與強拉電子基團(如 )構成推拉電子(Donor--Acceptor, )共軛體系,能顯著提升 躍遷的莫耳吸光係數()與自發放光速率常數 。
解題方法
由原題附圖可識別化合物 A 與 B 的關鍵結構特徵:
- 化合物 A:中央為雙氰基取代之碳-碳雙鍵(),兩端各連接一組帶有鄰位羥基()與對位二乙基胺基()的水楊醛亞胺單元。
- 化合物 B(水楊醛連氮,salicylaldazine):中央核心為單鍵連接的連氮基團(),兩端苯環僅有鄰位羥基,未帶有二乙基胺基。
分析兩者在乙腈(acetonitrile)溶液中螢光強弱差異的切入步驟如下:
- 比較中心連接鍵的旋轉自由度:分析雙鍵與單鍵對分子剛性與內部轉換()的影響。
- 比較取代基與共軛推拉電子體系:分析二乙基胺基()及氰基()對躍遷機率與輻射放光速率常數 的提升。
- 比較溶劑環境對非輻射淬滅的影響:乙腈為極性非質子溶劑,在此溶劑中分子自由溶解,結構柔韌分子的扭轉淬滅尤為明顯。
選項分析
本題為非選擇題(簡答與論述題),針對影響化合物 A 與 B 螢光強度的結構機制進行逐項對比分析:
- 骨架剛性與非輻射衰減(Structural Rigidity vs. Non-radiative Decay):
- 化合物 B 螢光微弱:其核心為氮-氮單鍵( single bond)。單鍵的旋轉能壘極低,在乙腈溶液中可自由發生分子內旋轉與扭曲振動(torsional vibration/intramolecular rotation)。此劇烈的分子運動提供了極為高效的內部轉換通道( 极大),使得激發態能量迅速轉變為熱能耗散,導致螢光嚴重淬滅(化合物 B 屬於典型的聚集誘導發光 AIE 分子,在稀溶液中發光極弱)。
- 化合物 A 螢光強烈:其核心為碳-碳雙鍵( double bond),雙鍵受 鍵重疊限制無法自由旋轉;且分子內酚羥基()與亞胺氮原子形成牢固的分子內氫鍵(),將全分子鎖定在高度剛硬且共平面的構形中。結構剛性極大程度抑制了內部轉換(),大幅提高螢光量子產率 。
- 取代基與共軛效應(Substituent Effect and Conjugation):
- 化合物 A:苯環對位具有兩個強給電子助色團(auxochrome)——二乙基胺基(),且中心帶有兩個強拉電子基團——氰基(),構成對稱的 延伸共軛體系。強烈的分子內電荷轉移(ICT)特性增大了電子躍遷偶極矩與莫耳吸光係數(),顯著提高激發態的輻射躍遷速率常數 。